N/A Completed N=20

Comparison of SureCALL® EMG Labor Monitor® and Tocodynamometer to Intrauterine Pressure Catheter Measurement (IUPC) Measurement in Pregnant Women

Pregnancy - Labor Monitoring
Source: ClinicalTrials.gov NCT01282710 ↗
Enrolled (actual)
20
Serious AEs
0.0%
Results posted
Mar 2011
Primary outcomePrimary: Comparison of SureCALL® and TOCO Detection of Contraction Events, as Compared to the IUPC — -0.73; -1.74 Seconds — p=.04627

Summary

This measurement study was designed to evaluate the performance of SureCALL® EMG Labor Monitor® (SureCALL®) to the predicate Tocodynamometer device (TOCO) using the performance of an Intrauterine Pressure Catheter device (IUPC) as the "gold standard".

Outcome Measures

OutcomeResultp-value
PRIMARY
Comparison of SureCALL® and TOCO Detection of Contraction Events, as Compared to the IUPC
-0.73; -1.74 .04627 sig

Eligibility Criteria

Inclusion Criteria

  • Singleton Pregnancy
  • Indicated for IUPC
  • Gestational ages from 37 to 41 weeks
  • Informed consent required

Exclusion Criteria

  • Multifetal Pregnancy
  • Not Indicated for IUPC
  • Gestation ages below 37 or above 41 weeks
  • Informed consent not given