Evaluation of a Ascensia ONYX NEXT Investigational Blood Glucose Monitoring System
Source: ClinicalTrials.gov NCT02390167 ↗Enrolled (actual)
376
Serious AEs
0.0%
Results posted
Apr 2017
Primary outcomePrimary: Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method — 99.4 Percent of Results within 15mg/dL/15%
Summary
The purpose of this study was to determine if untrained subjects with diabetes and without diabetes could operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid glucose results.
Outcome Measures
| Outcome | Result | p-value |
|---|---|---|
| PRIMARY Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method |
99.4 | — |
| SECONDARY Percent of Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method |
97.2 | — |
| SECONDARY Percent of Subject Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff |
99.7 | — |
| SECONDARY Percent of Venous Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff |
100 | — |
| SECONDARY Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range |
99.5 | — |
| SECONDARY Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range |
99.7 | — |
| SECONDARY Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range |
96.7 | — |
| SECONDARY Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range |
98.6 | — |
| SECONDARY Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Tested By Study Staff |
99.7 | — |
| SECONDARY Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Tested By Study Staff |
100 | — |
| SECONDARY Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 12.5mg/dL (<100mg/dL) and Within +/- 12.5% (>=100 mg/dL) of Laboratory Glucose Method |
98.5 | — |
| SECONDARY Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS |
97.9; 98.2; 99.7; 97.3; 97.9; 97.6 | — |
| SECONDARY Percent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS |
98.1; 98.4; 99.7; 97.6; 98.1; 97.6 | — |
| SECONDARY Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose |
100; 99.70; 99.40; 99.69; 99.70; 99.40 | — |
| SECONDARY Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method |
99.7 | — |
| SECONDARY Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method |
97.0 | — |
Eligibility Criteria
Inclusion Criteria
- Males and females, 18 years of age and older
- Able to speak, read, and understand English
- Willing to complete all study procedures
Exclusion Criteria
- Hemophilia or any other bleeding disorder
- Pregnancy
- Physical, visual or neurological impairments that would make the person unable to perform testing with the BGM
- Previously participated in a blood glucose monitoring study using the ONYX NEXT or ONYX PLUS BGMS
- Working for a medical laboratory, hospital, or other clinical setting that involves training on or clinical use of blood glucose monitors
- Working for a competitive medical device company, or having an immediate family member who works for such a company
- A condition which, in the opinion of the investigator or designee, would put the person or study conduct at risk
General enrollment guidelines:
- At least 70% of subjects will be younger than age 65
- At least 10% (approximately 10-15%) of subjects will be naive users
- At least 20% of subjects with diabetes will have type 1 diabetes
Data sourced from ClinicalTrials.gov (NCT02390167). Outcome figures and adverse-event rates are extracted automatically from the registry's posted results and are provided for clinician reference, not as a substitute for the primary publication. Informational only — not medical advice.