Mode
Text Size
Log in / Sign up
N/A Completed N=90 Diagnostic

Evaluation of the Accuracy of an Implanted Glucose Sensor

Source: ClinicalTrials.gov NCT02647905 ↗
Enrolled (actual)
90
Serious AEs
1.1%
Results posted
Jun 2018
Primary outcomePrimary: CGM Relative Difference to Laboratory Reference Reported as MARD — 8.5 percent

Summary

The purpose of this clinical investigation is to evaluate the accuracy of the Senseonics Continuous Glucose Monitoring System (Senseonics CGM System) measurements when compared with reference standard measurements The investigation will also evaluate safety of the Senseonics CGM System usage.

Outcome Measures

OutcomeResultp-value
PRIMARY
CGM Relative Difference to Laboratory Reference Reported as MARD
8.5 —

Eligibility Criteria

Inclusion Criteria

  • Adult subjects, age ≥18 years
  • Clinically confirmed diagnosis of diabetes mellitus for ≥1 year
  • Subject has signed an informed consent form and is willing to comply with protocol requirements

Exclusion Criteria

  • History of severe hypoglycemia in the previous 6 months. Severe hypoglycemia is defined as hypoglycemia resulting in loss of consciousness or seizure
  • History of diabetic ketoacidosis requiring emergency room visit or hospitalization in the previous 6 months
  • Female subjects of childbearing capacity (defined as not surgically sterile or not menopausal for ≥ 1 year) who are lactating or pregnant, intending to become pregnant, or not practicing birth control during the course of the study.
  • A condition preventing or complicating the placement, operation, or removal of the Sensor or wearing of transmitter, including upper extremity deformities or skin condition.
  • Symptomatic coronary artery disease; unstable angina; myocardial infarction, transient ischemic attack or stroke in the past 6 months; uncontrolled hypertension (systolic>160 mm Hg or diastolic >100 mm Hg at time of screening); current congestive heart failure; history of cardiac arrhythmia (benign PACs and PVCs allowed). Subjects with asymptomatic coronary artery disease (e,g, CABG, stent placement or angioplasty) may participate if negative stress test within 1 year prior to screening and written clearance from Cardiologist documented.
  • Hematocrit 55%
  • History of hepatitis B, hepatitis C, or HIV
  • Any condition that in the investigator's opinion would make the subject unable to complete the study or would make it not in the subject's best interest to participate in the study.
View full record on ClinicalTrials.gov →

Data sourced from ClinicalTrials.gov (NCT02647905). Outcome figures and adverse-event rates are extracted automatically from the registry's posted results and are provided for clinician reference, not as a substitute for the primary publication. Informational only — not medical advice.

Back to search