N/A Completed N=23 Other

Evaluation of Cochlear Baha 5 SuperPower Sound Processor on the Baha Attract System

Single Sided Sensorineural Deafness · Mixed Hearing Loss, Unilateral
Source: ClinicalTrials.gov NCT02722330 ↗
Enrolled (actual)
23
Serious AEs
0.0%
Results posted
Jun 2019
Primary outcomePrimary: Audiometry; Free Field Thresholds. Pure Tone Average PTA4,Unaided vs Aided — -27.7 dB HL — p=<0.0001

Summary

The aim of this clinical investigation is to evaluate objective and subjective hearing performance with the Cochlear Baha 5 SuperPower Sound Processor on the Baha Attract System, compared to the unaided hearing performance. Comparison of audiological test results with the same sound processor on a Baha Softband will also be performed. Short term safety parameters will be collected.

Outcome Measures

OutcomeResultp-value
PRIMARY
Audiometry; Free Field Thresholds. Pure Tone Average PTA4,Unaided vs Aided
-27.7 <0.0001 sig
SECONDARY
Change of the Self-reported Assessments of Hearing Outcome Using Abbreviated Profile of Hearing Aid Benefit (APHAB) From the Unaided Hearing Situation at Pre-operation With the Investigational Device on the Attract System, at 12 Weeks.
44.8; 31.5; 37.7; -16.6; 38.0 <0.0001 sig
SECONDARY
Change of the Self-reported Assessments of Hearing Outcome Using Speech, Spatial and Qualities of Hearing Scale (SSQ-12) From the Unaided Hearing Situation at Pre-operation With the Investigational Device on the Attract System, Aided at 12 Weeks.
3.59; 2.37; 3.51 <0.0001 sig
SECONDARY
Audiometry; Free Field Thresholds. Pure Tone Average PTA4 Unaided vs Aided
-27.0 <0.0001 sig
SECONDARY
Audiometry; Free Field Thresholds Per Frequency, Unaided vs Aided
-24.3; -27.6; -33.0; -32.2; -35.9; -32.4 <0.0001 sig
SECONDARY
Free Field; - Adaptive Speech Recognition in Noise, Unaided vs Aided
-8.41 0.0024 sig
SECONDARY
Free Field; - Speech Recognition in Quiet, Unaided vs Aided
35.7; 49.0; 39.2 <0.0001 sig
SECONDARY
Audiometry; Free Field Thresholds. Pure Tone Average PTA4, Investigational Device vs Softband
2.55 0.036 sig
SECONDARY
Audiometry; Free Field Thresholds. Pure Tone Average PTA4 [Mean of 0.5, 1, 2 and 4 kHz], Investigational Device vs Softband
1.9 0.065
SECONDARY
Audiometry; Free Field Thresholds Per Frequency, Investigational Device vs Softband
-2.39; -2.61; 1.09; 2.39; 1.52; 2.17 0.12
SECONDARY
Free Field - Adaptive Speech Recognition in Noise, Investigational Device vs Softband
1.42 0.11
SECONDARY
Free Field - Speech Recognition in Quiet, Investigational Device vs Softband
-9.00; -1.17; 2.82 0.096
SECONDARY
Sound Processor Magnet Selection at 4 Weeks
0; 2; 3; 11; 5; 2 —
SECONDARY
Sound Processor Magnet Selection at 6 Weeks
0; 5; 8; 7; 2; 1 —
SECONDARY
Sound Processor Magnet Selection at 12 Weeks
2; 3; 8; 10; 0; 0 —
SECONDARY
Daily Use of Sound Processor (Average Daily Use During the Last Week)
9.39 —

Eligibility Criteria

Inclusion Criteria

  • Adult subjects ≥ 18 years of age
  • Mixed hearing loss in the ear to be implanted. Bone conduction thresholds with pure tone average (PTA4, Mean of 0.5, 1, 2 and 4kHz) of 30-55 dBHL or
  • Single-sided sensorineural deafness: Bone conduction thresholds with a pure tone average [PTA4, Mean of 0.5, 1, 2 and 4 kHz] of lower or 30 dBHL in the best ear. Subjects who have, for some reason, experienced no or limited benefit from other sound processors when tested on a Baha Softband.
  • Subjects suitable for surgery according to the recommended surgical procedure for Cochlear Baha 5 SuperPower with no previous implant on the side to be investigated.
  • Signed inform consent.

Exclusion Criteria

  • Subject scheduled for simultaneous bilateral implant surgery. The investigation is limited to subjects with unilateral use of the Investigational device (however, bilateral hearing loss is not an exclusion criterion).
  • Suitable implant position for the BI300 Implant (4 mm or 3 mm) not found during surgery due to insufficient bone quality and/or bone thickness
  • Less than 3 mm soft tissue thickness at the planned implant site. (Intra-operative exclusion criteria judged by the surgeon at Visit 1 or Visit 2 according to local practice).
  • Subjects that have received radiation therapy at the same side of the skull where the Investigational device will be positioned.
  • Uncontrolled diabetes as judged by the investigator.
  • Condition that could jeopardise osseointegration and/or wound healing (e.g. osteoporosis, psoriasis and long-term systemic use of corticosteroids) or condition that may have an impact on the outcome of the investigation as judged by the investigator.
  • Unable to follow investigational procedures, e.g. to complete quality of life scales.
  • Participation in another clinical investigation with pharmaceutical and/or device.
View full record on ClinicalTrials.gov →