Mode
Text Size
Log in / Sign up
N/A Completed N=77

BioMonitor 2 In-Office Setting Insertion Safety and Feasibility Evaluation With Device Functionality Assessment

Insertable Cardiac Monitor
Source: ClinicalTrials.gov NCT02756338 ↗
Enrolled (actual)
77
Serious AEs
0.0%
Results posted
Sep 2018
Primary outcomePrimary: Number of Participants With Insertion Procedure-related Adverse Event That Requires Additional Invasive Intervention to Resolve — 0 participants with a Primary Objective AE

Summary

The purpose of this study is to evaluate the safety and feasibility of performing the BioMonitor 2 insertion procedure in an office setting.

Outcome Measures

OutcomeResultp-value
PRIMARY
Number of Participants With Insertion Procedure-related Adverse Event That Requires Additional Invasive Intervention to Resolve
— —
SECONDARY
Number of Participants With Insertion Procedure-related Adverse Event Not Included in Primary Objective
2 —
SECONDARY
Characterization of Insertion Procedure: Device Orientation
35; 38; 1; 3 —
SECONDARY
Characterization of Insertion Procedure: Incision Size
14.9 —
SECONDARY
Characterization of Insertion Procedure: Procedure Duration
8.4 —
SECONDARY
Characterization of Device Functionality Post-insertion: R-wave Amplitudes
0.77; 0.69; 0.67; 0.68 —

Eligibility Criteria

Inclusion Criteria

  • Indicated for continuous monitoring with an insertable cardiac monitor
  • Willing to be implanted in an office setting with only local anesthetic available
  • Able to understand the nature of the study and provide informed consent
  • Able and willing to complete all routine follow-up visits at the study site for the expected 90-day follow-up
  • Able and willing to use a CardioMessenger® capable of communicating with the BioMonitor 2
  • Age greater than or equal to 18 years

At the time of insertion, the following pre-procedure criteria must be met for the subject to undergo insertion:

  • Most recent INR value (within 7 days) is less than 3.5 if currently taking warfarin
  • Absence of infection with no history of infection within the last 30 days

Exclusion Criteria

  • Compromised immune system or at high risk of developing an infection
  • Abnormal thoracic anatomy or scar tissue at the implant site that may adversely impact the insertion procedure
  • Enrolled in any investigational cardiac device trial
  • Currently indicated for or implanted with a pacemaker, ICD device, or hemodynamic monitoring system
  • Currently implanted with an ICM or ILR
  • Life expectancy less than 6 months
  • Patients reporting pregnancy at the time of enrollment
View full record on ClinicalTrials.gov →

Data sourced from ClinicalTrials.gov (NCT02756338). Outcome figures and adverse-event rates are extracted automatically from the registry's posted results and are provided for clinician reference, not as a substitute for the primary publication. Informational only — not medical advice.

Back to search