Mode
Text Size
Log in / Sign up
N/A N=40

Evaluation of the Rectal Reflex Using Anal Acoustic Reflectometry

Pelvic Floor Dysfunction

Enrolled (actual)
40
Serious AEs
0.0%
Results posted
Mar 2020
Primary outcome: Primary: Opening Pressure — 34.44; 15.41 cmH2O

Study Design & Population

Study type
Observational
Phase
N/A
Interventions
Age
Adult, Older Adult · 18+ yrs
Sex
All
Sponsor
Manchester University NHS Foundation Trust
Primary completion
Oct 2016

Outcome Measures

OutcomeResultp-value
PRIMARY
Opening Pressure
34.44; 15.41
PRIMARY
Opening Elastance
1.39; 1.1
PRIMARY
Closing Pressure
20.06; 8.00
PRIMARY
Closing Elastance
1.31; 1.13
PRIMARY
Hysteresis
39.71; 39.9

Summary

Anal Acoustic Reflectometry (AAR) is a reliable and reproducible technique that has been studied in our department over the last 6 years. Sound waves pass into a balloon placed in the anal canal and are used to measure the cross-sectional area. By gradually increasing and decreasing the pressure in the balloon the investigators can measure the pressure at which the cross-sectional area starts to increase and decrease, and the anal canal starts to open and close. This assessment mimics the natural opening and closing of the anal canal and the effect of squeezing the muscles. The gold standard investigation of the anal sphincter muscles has been manometry which measures anal canal pressure at rest and during squeeze. However, it has limitations. In previous studies AAR has shown promise in the assessment of faecal incontinence and, that unlike manometry, it has been able to distinguish between different types of incontinence. The Recto-anal Inhibitory Reflex (RAIR) is a normal response when the rectum fills with faeces, fluid or air, whereby there is a change in the pressures within the anal canal to determine the type of contents. This can be absent or altered in patients who have difficulty in opening their bowels. The RAIR is currently measured by anal manometry using a 4.9mm catheter, resulting in an anal canal which is already partially opened prior to the measurement, and potentially distorted. AAR is considered a catheter free technique as the balloon has a cross-sectional area of only 0.4mm2 when collapsed. The investigators propose to measure the RAIR using a 1.7mm diameter catheter alongside the AAR balloon to determine the effect that its placement has on the recorded parameters of AAR. This aims to improve our understanding of the opening and closing of the anal canal in response to distension of the rectum.

Eligibility Criteria

Inclusion Criteria

  • Adults over 18 years old
  • Have capacity to consent to the study
  • Patients with pelvic floor dysfunction.

Exclusion Criteria

  • Minors under the age of 18 years old
  • Patients who lack capacity to consent
  • Patients with known rectal tumours and proctitis
  • Asymptomatic patients
View full record on ClinicalTrials.gov →

Data sourced from ClinicalTrials.gov (NCT02782377). Outcome figures and adverse-event rates are extracted automatically from the registry's posted results and are provided for clinician reference, not as a substitute for the primary publication.