Mode
Text Size
Log in / Sign up
N/A N=50 Diagnostic

Dario Blood Glucose Monitoring System - Android Devices Altitude Evaluation Study

Diabetes Mellitus

Enrolled (actual)
50
Serious AEs
0.0%
Results posted
Oct 2020
Primary outcome: Primary: Number of Successful Glucose Readings — 449 Number of Glucose readings in range

Study Design & Population

Study type
Interventional
Phase
N/A
Interventions
Blood Glucose Monitoring System Altitude Performance (Device)
Age
Adult, Older Adult · 18+ yrs
Sex
All
Sponsor
LabStyle Innovations Ltd.
Primary completion
Jul 2016

Outcome Measures

OutcomeResultp-value
PRIMARY
Number of Successful Glucose Readings
449

Summary

This study evaluates the performance of a blood glucose monitoring system at high altitudes (approximately 10,000 feet). The performance is assessed via a reference method for glucose analysis (Yellow Springs Instrument 2300 STAT Plus Glucose Analyzer).

Eligibility Criteria

Inclusion Criteria

  • Able to speak and read English

Exclusion Criteria

  • Pregnant
  • Has any condition that the principle investigator believes may interfere in the subject's participation in the trial.
View full record on ClinicalTrials.gov →

Data sourced from ClinicalTrials.gov (NCT02916563). Outcome figures and adverse-event rates are extracted automatically from the registry's posted results and are provided for clinician reference, not as a substitute for the primary publication.