Mode
Text Size
Log in / Sign up
N/A Completed N=118 Other

Accuracy of Hemoglobin Measurement for Various Rainbow Pulse Oximeter Sensors

Healthy
Source: ClinicalTrials.gov NCT03124966 ↗
Enrolled (actual)
118
Serious AEs
0.0%
Results posted
Jun 2017
Primary outcomePrimary: Accuracy of Sensor by Arms Calculation — 0.96 g/dL

Summary

In this study, the concentration of hemoglobin within the subject's blood will be reduced in a controlled manner by administering fluids intravenously. The accuracy of a noninvasive hemoglobin sensor(s) will be assessed by comparison to hemoglobin measurements from a laboratory analyzer.

Outcome Measures

OutcomeResultp-value
PRIMARY
Accuracy of Sensor by Arms Calculation
0.96 —

Eligibility Criteria

Study Site 1:

Inclusion Criteria

  • Has physical status between ASA 1 or 2
  • Able to communicate in English

Exclusion Criteria

  • Pregnant or sexually active without birth control.
  • Hemoglobin less than 11g/dL
  • Known alcohol or drug abuse
  • Skin abnormalities affecting the digits such as psoriasis, eczema, angioma, scar tissue, burn, fungal infection
  • Nail polish
  • Head injury with loss of consciousness within the last year
  • Known neurological and psychiatric conditions.
  • Known concurrent chronic usage of psychoactive or anti-convulsive drugs within the last 90 days, or use in the last 7 days (i.e. tricyclic antidepressants, MAO inhibitors, lithium, neuroleptics, anxiolytics or antipsychotics. Except SSRIs
  • Subject has any medical condition which, in the judgement of the investigator, renders them inappropriate for participation in this study, such as Raynaud's syndrome
  • Hypertension: Systolic BP >= 140 mmHg or Diastolic BP >= 90 mmHg
  • Baseline heart rate 140 mmHg or Diastolic BP > 90 mmHg.
  • Baseline heart rate < 50 bpm.
  • Inability to tolerate sitting still or minimal movement for up to 90 minutes
View full record on ClinicalTrials.gov →

Data sourced from ClinicalTrials.gov (NCT03124966). Outcome figures and adverse-event rates are extracted automatically from the registry's posted results and are provided for clinician reference, not as a substitute for the primary publication. Informational only — not medical advice.

Back to search