Mode
Text Size
Log in / Sign up
N/A N=203 Screening

BiliCam Clinical Validation Study

Newborn Jaundice

Enrolled (actual)
203
Serious AEs
0.0%
Results posted
Jun 2022
Primary outcome: Primary: BCB - TSB Correlation — 0.82 correlation coefficient — p=<.001

Study Design & Population

Study type
Interventional
Phase
N/A
Interventions
BiliCam estimated bilirubin (BCB) (Device)
Age
Pediatric · 0+ yrs
Sex
All
Sponsor
BiliCam, LLC
Primary completion
Feb 2018

Outcome Measures

OutcomeResultp-value
PRIMARY
BCB - TSB Correlation
0.82 <.001 sig
SECONDARY
Intercept of Regression Line
4.13 <.001 sig
SECONDARY
Slope of Regression Line
0.68 <.001 sig

Summary

The goal of this study is to assess the accuracy of BiliCam, a non-invasive technology based on analysis of digital images obtained with an app installed on a commercial smartphone, in estimating total serum bilirubin levels in newborns. BiliCam estimated bilirubin levels will be compared to total serum bilirubin levels in up to 225 newborns.

Eligibility Criteria

Inclusion Criteria

  • Newborn born at ≥ 35 weeks gestation
  • Parent speaks and reads English
  • Parent provides written informed consent

Exclusion Criteria

  • previous or ongoing treatment with phototherapy for hyperbilirubinemia
  • Medical or other complications that preclude completion of the study
View full record on ClinicalTrials.gov →

Data sourced from ClinicalTrials.gov (NCT03246503). Outcome figures and adverse-event rates are extracted automatically from the registry's posted results and are provided for clinician reference, not as a substitute for the primary publication.

Back to search