N/A Completed N=203 Screening

BiliCam Clinical Validation Study

Newborn Jaundice
Source: ClinicalTrials.gov NCT03246503 ↗
Enrolled (actual)
203
Serious AEs
0.0%
Results posted
Jun 2022
Primary outcomePrimary: BCB - TSB Correlation — 0.82 correlation coefficient — p=<.001

Summary

The goal of this study is to assess the accuracy of BiliCam, a non-invasive technology based on analysis of digital images obtained with an app installed on a commercial smartphone, in estimating total serum bilirubin levels in newborns. BiliCam estimated bilirubin levels will be compared to total serum bilirubin levels in up to 225 newborns.

Outcome Measures

OutcomeResultp-value
PRIMARY
BCB - TSB Correlation
0.82 <.001 sig
SECONDARY
Intercept of Regression Line
4.13 <.001 sig
SECONDARY
Slope of Regression Line
0.68 <.001 sig

Eligibility Criteria

Inclusion Criteria

  • Newborn born at ≥ 35 weeks gestation
  • Parent speaks and reads English
  • Parent provides written informed consent

Exclusion Criteria

  • previous or ongoing treatment with phototherapy for hyperbilirubinemia
  • Medical or other complications that preclude completion of the study
View full record on ClinicalTrials.gov →

Data sourced from ClinicalTrials.gov (NCT03246503). Outcome figures and adverse-event rates are extracted automatically from the registry's posted results and are provided for clinician reference, not as a substitute for the primary publication. Informational only — not medical advice.