Mode
Text Size
Log in / Sign up
N/A N=110

Clinical Accuracy of the welloStationX

Fever

Enrolled (actual)
110
Serious AEs
0.0%
Results posted
Jul 2021
Primary outcome: Primary: Clinical Agreement Through Clinical Bias Evaluation — 0.04 degrees Farehheit

Study Design & Population

Study type
Observational
Phase
N/A
Interventions
welloStationX (Diagnostic_test); SureTemp (Diagnostic_test)
Age
Pediatric, Adult, Older Adult · 5+ yrs
Sex
All
Sponsor
Wello Inc.
Primary completion
Nov 2018

Outcome Measures

OutcomeResultp-value
PRIMARY
Clinical Agreement Through Clinical Bias Evaluation
0.04
SECONDARY
Clinical Repeatability
0.062

Summary

This study evaluates the clinical accuracy of the welloStationX, an automated non-touch self-service electronic thermometer using an infrared sensor of the surface of the forehead to screen for fever.

Eligibility Criteria

Inclusion Criteria

  • Male or female.
  • 5 years and up.
  • Signed informed consent form.

Exclusion Criteria

  • Use of medications known to affect body temperature (for example, antipyretics, barbiturates, thyroid preparations, antipsychotics, etc.) within 3 hours of the test.
  • immunization within seven days of the test
View full record on ClinicalTrials.gov →

Data sourced from ClinicalTrials.gov (NCT03758157). Outcome figures and adverse-event rates are extracted automatically from the registry's posted results and are provided for clinician reference, not as a substitute for the primary publication.

Back to search