Completed N=108
Comparison of Depth of Sedation Performance Between SedLine and Comparator Device During General Anesthesia
Anesthesia
Source: ClinicalTrials.gov NCT03865316 ↗
Enrolled (actual)
108
Serious AEs
0.0%
Results posted
Apr 2022
Primary outcomePrimary: Correlation of Sedation Indices Performance Across Anesthesia Range — 0.8314 Pearson r
Summary
The primary objective is to compare the performance of Masimo's SedLine and the comparator's device systems during surgery. Relative accuracy of the individual depth of sedation indices will be compared.
Outcome Measures
| Outcome | Result | p-value |
|---|---|---|
| PRIMARY Correlation of Sedation Indices Performance Across Anesthesia Range |
0.8314 | — |
Eligibility Criteria
Inclusion Criteria
- Patients 18 years old and older at the time of consent.
- ASA status I, II, or III.
- English-speaking subjects.
- Scheduled for surgical and non-surgical procedures scheduled under general anesthesia (Common procedures include but are not restricted to tonsillectomy, adenoidectomy, urological procedures, dental rehabilitation, orthopedic procedures, biopsies, audiogram, nuclear scans, gastroscopy, colonoscopy, etc.).
Exclusion Criteria
- Any deformities or devices that may prevent application of EEG Sensor to forehead with a proper fit.
- Subjects who are developmentally delayed.
- Subjects deemed not suitable for study at the discretion of the investigator.