N/A N=30 Randomized Treatment

Implant Primary Stability and Ridge Dimensional Changes Utilizing the Osseodensification Protocol

Dental Implants

Enrolled (actual)
30
Serious AEs
0.0%
Results posted
Nov 2024
Primary outcome: Primary: Change in Bone Ridge Width at Crest — 0.84; 0.01 millimeters

Study Design & Population

Study type
Interventional
Phase
N/A
Interventions
OsseoDensification (OD) protocol (Device); standard drilling (SD) protocol (NobelReplace® Conical) (Device)
Age
Adult, Older Adult · 22+ yrs
Sex
All
Sponsor
The University of Texas Health Science Center, Houston
Primary completion
Nov 2023

Outcome Measures

OutcomeResultp-value
PRIMARY
Change in Bone Ridge Width at Crest
0.84; 0.01
PRIMARY
Change in Bone Ridge Width at 5mm Apical From Crest
0.62; 0.11
PRIMARY
Change in Bone Ridge Width at 10mm Apical From Crest
0.29; 0.01
SECONDARY
Primary Stability of the Implants as Measured by the Insertion Torque Values
57.7; 45
SECONDARY
Primary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter Baseline(After Implant Placement)
73.7; 67.0
SECONDARY
Primary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter 3 Weeks After Implant Placement
75; 69.8
SECONDARY
Primary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter 6 Weeks After Implant Placement
75.3; 70.8
SECONDARY
Primary Stability of the Implants as Indicated by the Implant Stability Quotient (ISQ) as Measured by the Resonance Frequency Analysis Using the Ostell ISQ Meter 12 Weeks After Implant Placement
76.8; 73.3
SECONDARY
Volume of the Ridge as Assessed by the Cone Beam Computer Tomography (CBCT)
SECONDARY
Percent Change of Buccal Bone Thickness as Assessed by the Cone Beam Computer Tomography at 1.5 mm From the Implant Platform
6.24; -21.3
SECONDARY
Percent Change of Buccal Bone Thickness as Assessed by the Cone Beam Computer Tomography at 5 mm From the Implant Platform
8.61; -15.1

Summary

The purpose of this study is to compare ridge dimensional changes between densification and standard drilling protocols and to compare primary and secondary stability of implants placed by bone densification and standard drilling protocols

Eligibility Criteria

Inclusion Criteria

  • Patients who understand and agree to this study.
  • Adequate Oral hygiene
  • One or more edentulous spaces, which are 10 weeks or longer after extractions

Exclusion Criteria

  • Pregnancy.
  • Smoking of more than 5 cigarettes/day.
  • History of alcoholism or drug abuse during the last 5 years.
  • Uncontrolled Hypertension or diabetes.
  • Patient with malignant tumor.
  • Patients on daily dose of steroids.
  • Patients with history of chemotherapy or radiation for the last 12 months.
View full record on ClinicalTrials.gov →