Mode
Text Size
Log in / Sign up
N/A Completed N=120

Performance of Contour Next® and Contour Plus Elite® BGMS in Arterial Blood Samples From Hospitalized Adults

Source: ClinicalTrials.gov NCT05449795 ↗
Enrolled (actual)
120
Serious AEs
0.0%
Results posted
Jan 2025
Primary outcomePrimary: Number of Results From the Contour Next BGMSs Reference Values Within ±12.5% of Reference Values — 236 blood samples

Summary

The purpose of the study is to extend the intended use of two BGMSs to include testing of arterial blood by Health Care Professionals (HCP) in a clinical setting.

Outcome Measures

OutcomeResultp-value
PRIMARY
Number of Results From the Contour Next BGMSs Reference Values Within ±12.5% of Reference Values
236
PRIMARY
Number of Results From the Contour Plus Elite BGMSs Reference Values Within ±12.5% of Reference Values
240

Eligibility Criteria

Inclusion Criteria

  • Patients who are at least 18 years old.
  • Residual Arterial blood samples collected from subjects hospitalized in a Critical Care Unit (medical and surgical intensive care units (ICUs)).
  • Sample blood volume must be sufficient to complete investigational testing procedures clinical laboratory testing.

Exclusion Criteria

  • Residual arterial blood samples collected from subjects previously enrolled. and evaluated for this study.
View full record on ClinicalTrials.gov →

Data sourced from ClinicalTrials.gov (NCT05449795). Outcome figures and adverse-event rates are extracted automatically from the registry's posted results and are provided for clinician reference, not as a substitute for the primary publication. Informational only — not medical advice.

Back to search