N/A N=86 Other

Clinical Performance of Masimo INVSENSOR00057 for Heart Rate Measurements

Healthy

Enrolled (actual)
86
Serious AEs
0.0%
Results posted
May 2024
Primary outcome: Primary: Heart Rate Accuracy of Masimo INVSENSOR00057 — 1.38 bpm

Study Design & Population

Study type
Interventional
Phase
N/A
Interventions
Masimo INVSENSOR00057 (Device)
Age
Adult, Older Adult · 18+ yrs
Sex
All
Sponsor
Masimo Corporation
Primary completion
Jul 2022

Outcome Measures

OutcomeResultp-value
PRIMARY
Heart Rate Accuracy of Masimo INVSENSOR00057
1.38

Summary

This study is designed to compare the accuracy of a noninvasive measurement of heart rate compared to reference values obtained by a standard of care ECG monitor.

Eligibility Criteria

Inclusion Criteria

  • Subject is 18 to 80 years of age.
  • Subject is able to read and communicate in English.

Exclusion Criteria

  • Subject has open wounds, inflamed tattoos or piercings on the area of device placement, and/or has any visible healing wounds that the investigator and/or medical professional determines may place them at an increased risk for participation.
  • Subject has any medical condition which in the judgment of the investigator and/or study staff, renders them ineligible for participation in this study or subject is deemed ineligible by the discretion of the investigator/study staff.
View full record on ClinicalTrials.gov →