Mode
Text Size
Log in / Sign up
N/A Completed N=18 Other

Dexcom G6 for Total Pancreatectomy With Islet Autotransplantation (TPIAT) Patients

Source: ClinicalTrials.gov NCT05607576 ↗
Enrolled (actual)
18
Serious AEs
0.0%
Results posted
May 2025
Primary outcomePrimary: Dexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %) — 806; 1226; 704; 1715 Matched blood glucose readings — p=<0.05

Summary

A single-center prospective trial on the effects of standard of care radiation exposure on the Dexcom G6 continuous glucose monitoring (CGM) device.

Outcome Measures

OutcomeResultp-value
PRIMARY
Dexcom G6 Accuracy Post Radiation Exposure by Clark Error Analysis (Zone A and B %)
806; 1226; 704; 1715; 1020; 606 <0.05 sig
PRIMARY
Post Radiation Dexcom G6 Transmitter Functional Duration by Clark Error Analysis (Zone A and B %)
2065; 446; 226; 1173; 128; 61 <0.05 sig

Eligibility Criteria

Inclusion Criteria

  • Planned TPIAT at CCHMC
  • Staying in-patient at CCHMC post TPIAT
  • Planned to receive a Dexcom G6 CGM

Exclusion Criteria

  • Hemo- or peritoneal dialysis
View full record on ClinicalTrials.gov →

Data sourced from ClinicalTrials.gov (NCT05607576). Outcome figures and adverse-event rates are extracted automatically from the registry's posted results and are provided for clinician reference, not as a substitute for the primary publication. Informational only — not medical advice.