Calibration and Validation of the Masimo Temperature Device in Febrile Patients
Fever
Bottom Line
View on ClinicalTrials.gov: NCT05674344 ↗Enrolled (actual)
170
Serious AEs
0.0%
Results posted
Jul 2023
Primary outcome: Primary: Accuracy of Masimo RAD-GT Temperature Measurments — -0.24; 0.85; 0.09 Degrees Celsius
Study Design & Population
- Study type
- Interventional
- Phase
- N/A
- Interventions
- Masimo Temperature device (Device)
- Age
- Pediatric, Adult, Older Adult · 1+ yrs
- Sex
- All
- Sponsor
- Masimo Corporation
- Primary completion
- Mar 2023
Outcome Measures
| Outcome | Result | p-value |
|---|---|---|
| PRIMARY Accuracy of Masimo RAD-GT Temperature Measurments |
-0.24; 0.85; 0.09 | — |
Summary
The goal of this clinical trial is to test the performance of the Masimo noninvasive temperature device in patients with fever. The main question it aims to answer is whether the Masimo temperature device is comparable to standard methods of taking temperature such as inserting a probe under the tongue or using a forehead thermometer.
Participants will be asked to sit still while rounds of measurements are taken using the Masimo temperature device and the reference temperature device. Researchers will compare the measurements taken with the Masimo device with the measurements taken with the reference device.
Eligibility Criteria
Inclusion Criteria
- Patients age ≥ 1year old
- Febrile at time of enrollment: ≥ 99.5oF/37.5oC for patients ages 1-5 years; ≥ 100.4oF/38oC for patients older than 5 years old.
- English- or Spanish-speaking patient or parent/LAR
Exclusion Criteria
- Patients deemed not suitable for the study at the discretion of the investigator
- Patients who are rated as a 1 or 2 Emergency Severity Index
Data sourced from ClinicalTrials.gov (NCT05674344). Outcome figures and adverse-event rates are extracted automatically from the registry's posted results and are provided for clinician reference, not as a substitute for the primary publication.