Calibration and Validation of the Masimo Temperature Device in Febrile Patients

Fever

Enrolled (actual)
170
Serious AEs
0.0%
Results posted
Jul 2023
Primary outcome: Primary: Accuracy of Masimo RAD-GT Temperature Measurments — -0.24; 0.85; 0.09 Degrees Celsius

Study Design & Population

Study type
Interventional
Phase
N/A
Interventions
Masimo Temperature device (Device)
Age
Pediatric, Adult, Older Adult · 1+ yrs
Sex
All
Sponsor
Masimo Corporation
Primary completion
Mar 2023

Outcome Measures

OutcomeResultp-value
PRIMARY
Accuracy of Masimo RAD-GT Temperature Measurments
-0.24; 0.85; 0.09

Summary

The goal of this clinical trial is to test the performance of the Masimo noninvasive temperature device in patients with fever. The main question it aims to answer is whether the Masimo temperature device is comparable to standard methods of taking temperature such as inserting a probe under the tongue or using a forehead thermometer. Participants will be asked to sit still while rounds of measurements are taken using the Masimo temperature device and the reference temperature device. Researchers will compare the measurements taken with the Masimo device with the measurements taken with the reference device.

Eligibility Criteria

Inclusion Criteria

  • Patients age ≥ 1year old
  • Febrile at time of enrollment: ≥ 99.5oF/37.5oC for patients ages 1-5 years; ≥ 100.4oF/38oC for patients older than 5 years old.
  • English- or Spanish-speaking patient or parent/LAR

Exclusion Criteria

  • Patients deemed not suitable for the study at the discretion of the investigator
  • Patients who are rated as a 1 or 2 Emergency Severity Index
View full record on ClinicalTrials.gov →