Validation of the Masimo INVSENSOR00063 in Febrile and Afebrile Subejcts

Fever

Enrolled (actual)
60
Serious AEs
0.0%
Results posted
Jul 2023
Primary outcome: Primary: Accuracy of Temperature Measurements Using Masimo INVSENSOR00063 — -0.24; 0.85; 0.05 Degrees Celsius

Study Design & Population

Study type
Interventional
Phase
N/A
Interventions
Masimo INVSENSOR00063 (Device)
Age
Pediatric, Adult, Older Adult · 1+ yrs
Sex
All
Sponsor
Masimo Corporation
Primary completion
Mar 2023

Outcome Measures

OutcomeResultp-value
PRIMARY
Accuracy of Temperature Measurements Using Masimo INVSENSOR00063
-0.24; 0.85; 0.05

Summary

This is a prospective, non-randomized data collection study to evaluate the performance of the Masimo INVSENSOR00063 in obtaining temperature measurements.

Eligibility Criteria

Inclusion Criteria

  • Participant age ≥ 1 year
  • Afebrile or febrile at time of enrollment: For febrile subjects, ≥ 99.5 F/ 37.5 C for sublingual temperature reference; ≥ 100.4 F/ 38 C for temporal artery temperature reference.
  • English-or Spanish-speaking patient or parent/Legal Guardian

Exclusion Criteria

  • Participants deemed not suitable for the study at the discretion of the investigator
View full record on ClinicalTrials.gov →