Mode
Text Size
Log in / Sign up

Clinical Trial Search

Completed trials with posted results — search by concept, filter by the parameters that matter at the bedside.

Phase
Intervention
Population
Results posted

30 closest matches · ranked by relevance

N/A N=229 ProMRI Study of the Entovis Pacemaker System

Safety of the ProMRI Pacemaker System Under Specific MRI Conditions
Primary: MRI and Pacing System Related Serious Adverse Device Effect (SADE) Free Rate — 100 percentage of participants
Biotronik, Inc. Results Nov 2014 9.2% serious AE View details

N/A N=154 ProMRI Study of the Iforia ICD System (Phase C)

Implantable Cardioverter-Defibrillator With Magnetic Resonance Imaging
Primary: MRI and ICD System Related Serious Adverse Device Effect (SADE) Free Rate — 100 percentage of participants
Biotronik, Inc. Results Oct 2015 14.9% serious AE View details

N/A N=283 Accent Cardiac MRI Study

Bradycardia
Primary: Freedom From MRI Scan-related Complications — 123 Participants — p=<0.01
Abbott Medical Devices Results Aug 2019 4.2% serious AE View details

N/A N=239 ENABLE MRI (Magnetic Resonance Imaging)

Magnetic Resonance Imaging
Primary: Percentage of Participants Free From MR (Magnetic Resonance) Scan-related Complications — 100 percentage of participants
Boston Scientific Corporation Results Jul 2019 28.1% serious AE View details

N/A N=464 EMRI SureScan™ Clinical Study

Bradycardia · Slow Heart Beat
Primary: Magnetic Resonance Imaging (MRI)-Related Complications — 0 participants — p=<0.001
Medtronic Cardiac Rhythm and Heart Failure Results Aug 2011 40.5% serious AE View details

N/A N=263 Advisa MRI Clinical Study

Magnetic Resonance Imaging · Cardiac Pacemaker, Artificial
Primary: Magnetic Resonance Imaging (MRI)-Related Complications — 0 participants — p=<0.0001
Medtronic Cardiac Rhythm and Heart Failure Results Apr 2013 25.1% serious AE View details

N/A N=63 China 3T MRI Study

Bradycardia · Tachycardia · Heart Failure
Primary: The Occurrence of an MRI-related Event for CIED — 59; 0 Participants — p=0.002
Medtronic Cardiac Rhythm and Heart Failure Results Jan 2026 7.8% serious AE View details

N/A N=725 Micra Transcatheter Pacing Study

Class I or II Indication for Implantation of a Single Chamber Ventricular Pacemaker According to ACC/AHA/HRS 2001 Guidelines and Any National Guidelines
Primary: Major Complications — 96.0 Kaplan-Meier survival probability (%) — p=<0.0001
Medtronic Cardiac Rhythm and Heart Failure Results Jan 2017 31.2% serious AE View details

N/A N=189 SJM MRI Diagnostic Imaging Registry (IDE)

Cardiac Arrhythmia
Primary: The Proportion of MRI Scans From Pacemakers or CRT-Ps Providing Sufficient Image Quality to Allow for a Diagnostic Interpretation. — 129 MRI Scans
Abbott Medical Devices Results Aug 2019 0.0% serious AE View details