Safety and Efficacy Study of the QuickFlex Micro Model 1258T Left Heart Pacing Lead
Source: ClinicalTrials.gov NCT00763698 ↗Enrolled (actual)
86
Serious AEs
4.7%
Results posted
Nov 2011
Primary outcomePrimary: Freedom From Left Ventricular Lead-related Complications — 100 percent of participants
Summary
This is a prospective, multi-center non-randomized clinical study to evaluate the safety and efficacy of the QuickFlex Micro Model 1258T left ventricular lead in a heart failure patient population. The study will be conducted at a maximum of 20 investigational centers located in the United States and a total of 68 patients will be enrolled.
Outcome Measures
| Outcome | Result | p-value |
|---|---|---|
| PRIMARY Freedom From Left Ventricular Lead-related Complications |
100 | — |
| PRIMARY Percentage of Successful Left Ventricular Lead Implants |
94 | — |
| PRIMARY Left Ventricular Bipolar Pacing Capture Threshold (Volts) |
1.1 | — |
Eligibility Criteria
Inclusion Criteria
- Approved indication for CRT-D system
Exclusion Criteria
- Previous cerebrovascular accident (CVA) or transient ischemic attack (TIA) within 3 months of enrollment
- Previous left ventricular (LV) lead implant
- Have had a heart attack, unstable angina or cardiac revascularization (percutaneous transluminal coronary angioplasty (PTCA),coronary artery bypass grafting (CABG), Stent) within 40 days prior to enrollment