Mode
Text Size
Log in / Sign up
N/A Completed N=458 Randomized Double-blind Treatment

Randomized MicroPort's Firehawk DES Versus Xience V: TARGET I Trial

Source: ClinicalTrials.gov NCT01196819 ↗
Enrolled (actual)
458
Serious AEs
16.4%
Results posted
Feb 2018
Primary outcomePrimary: 9 Months In-stent Late Lumen Loss — 0.13; 0.13 mm — p=0.94

Summary

The purpose of this study is to evaluate new generation drug eluting stent (DES) of MicroPort of its safety, efficacy and delivery system in treating CAD.

Outcome Measures

OutcomeResultp-value
PRIMARY
9 Months In-stent Late Lumen Loss
0.13; 0.13 0.94
SECONDARY
9 Months In-stent Diameter Stenosis
11.51; 11.85 0.69
SECONDARY
Target Lesion Failure(TLF) Rate
2.2; 2.2 1.0
SECONDARY
TLF(Target Lumen Failure) Rate
3.51; 4.05 0.76
SECONDARY
TLF(Target Lumen Failure)
6.67; 5.99 0.77
SECONDARY
Number of Participants With Stent Thrombsis (ARC Defined Definite/Probable)
1; 0 —

Eligibility Criteria

Inclusion Criteria

  • Age 18-75, male or women who are not pregnant
  • Evidence of non-symptomatic ischemia, stable or non-stable angina or past heart attack cases
  • Target lesion is primary, single artery and single lesion of coronary artery
  • Target lesion vessel length ≤24mm, diameter 2.25mm-4.0mm
  • Lesion diameter stenosis ≥70%
  • Candidates understand the study, willing to sign Consent of Agreement and to willing to accept follow-up For long lesion group, at least one target lesion length ≥28mm,diameter 2.5mm-4.0mm, and one lesion needs to implant 33mm 0r 38mm long stent

Exclusion Criteria

  • Acute heart attack within one week
  • Chronic complete stenosis (TIMI 0), left main lesion, three branches need to treat, branch vessel diameter ≤2.5mm and bypass lesion
  • Calcified lesion failed in pre-dilation and twisted lesion
  • In-stent restenosis
  • Stent implanted within one year
  • Severe heart failure (NYHA above III) or left ventricle EF 2.0mg/dl
  • Bleeding risk; allergic to drugs and agents used in procedure/treatment
  • Life expectation < 12 months
  • No compliances to the protocol
  • Heart implantation cases

Pre-specified OCT inclusion/exclusion criteria:

  • Patients underwent 9-month angio F/U
  • No binary restenosis at 9-month
  • LLL between -0.01~0.2mm
  • No mix-implanted stent
  • No more than one bail-out stent
  • No TLR occurred at 3 years after PCI
  • The target vessel able to deliver the OCT catheter
View full record on ClinicalTrials.gov →

Data sourced from ClinicalTrials.gov (NCT01196819). Outcome figures and adverse-event rates are extracted automatically from the registry's posted results and are provided for clinician reference, not as a substitute for the primary publication. Informational only — not medical advice.

Back to search