N/A Completed N=6,700

Adult and Paediatric Patient Radiation Doses From Multidetector Row Computed Tomography Scans: a National Survey

Source: ClinicalTrials.gov NCT01436006 ↗
Enrolled (actual)
6,700
Serious AEs
0.0%
Results posted
Aug 2015
Primary outcomePrimary: Value of Mean Volumetric Computed Tomography Dose Index (CTDI Vol) for Head CT Examinations. — 64; 28.7 mGy

Summary

The situation in Italy regarding radiation dose to patients undergoing Multidetector row Computed Tomography (MDCT) scans is unclear despite Italian law require that each centre check the Reference Dose Level (RDL) every two years. RDL in CT were first suggested in a document of 1998 published by the European Commission (EUR16262/1998) based on the results of an older survey in the UK from the late 1980s. Those RDLs, expressed in terms of Computed Tomography Dose Index (CTDI) and Dose Length Product (DLP), are reported in the Italian law (D.Lgs.187/00). However, in Italy there is no governmental authority comparable to the UK's National Radiological Protection Board, deputed to dose evaluation and protection. An overview of the Italian situation is relevant in order to know the actual dose levels for MDCT administered to patients both adults and paediatrics.

Outcome Measures

OutcomeResultp-value
PRIMARY
Value of Mean Volumetric Computed Tomography Dose Index (CTDI Vol) for Head CT Examinations.
64; 28.7
PRIMARY
Value of Mean Volumetric Computed Tomography Dose Index (CTDI Vol) for Chest CT Examinations.
12; 4.3
PRIMARY
Value of Mean Volumetric Computed Tomography Dose Index (CTDI Vol) for Abdomen CT Examinations.
15; 8.9
PRIMARY
Value of Mean Volumetric Computed Tomography Dose Index (CTDI Vol) for Cardiac CT Examinations.
43
PRIMARY
Value of Mean Volumetric Computed Tomography Dose Index (CTDI Vol) for Chest Abdomen and Pelvis CT Examinations.
14
PRIMARY
Value of Mean Volumetric Computed Tomography Dose Index (CTDI Vol) for Spine CT Examinations.
34

Eligibility Criteria

Inclusion Criteria

  • Patients that will be submitted to "standard" protocols for CT of the: head, chest, abdomen, lumbar spine and cardiac

Exclusion Criteria

  • Non standard MDCT examination
View full record on ClinicalTrials.gov →

Data sourced from ClinicalTrials.gov (NCT01436006). Outcome figures and adverse-event rates are extracted automatically from the registry's posted results and are provided for clinician reference, not as a substitute for the primary publication. Informational only — not medical advice.

Back to search