N/A Completed N=50

Study of Clinical Performance of Corneal Topography Measurement Using an Optical Biometer

Corneal Topography
Source: ClinicalTrials.gov NCT01861925 ↗
Enrolled (actual)
50
Serious AEs
0.0%
Results posted
Jun 2014
Primary outcomePrimary: In-vivo Repeatability of Corneal Topography Measurements With Lenstar LS 900 Topography: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences Between Two Consecutive Measurements — -0.014; 0.053; -0.120; -0.029 diopters

Summary

The purpose of this study is to quantify in-vivo repeatability of a new corneal topography measurement mode of an optical biometer, and the clinical performance of measurement in comparison to reference devices.

Outcome Measures

OutcomeResultp-value
PRIMARY
In-vivo Repeatability of Corneal Topography Measurements With Lenstar LS 900 Topography: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences Between Two Consecutive Measurements
-0.014; 0.053; -0.120; -0.029; 0.015; -0.089 —
SECONDARY
Equivalence of Corneal Topography Measurement Between Lenstar LS 900 Topography and Atlas 9000: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences for One Measurement Per Device.
0.044; 0.012; 0.160; 0.79; 0.77; 2.49 —

Eligibility Criteria

Inclusion Criteria

  • 18 years or older

Exclusion Criteria

  • No corneal transplants
  • No corneal scar(s)
  • No acute infections of the eye
  • No weak fixation
  • No suspected lack of compliance
  • No vulnerable population
View full record on ClinicalTrials.gov →