N/A N=11 Treatment

Pilot Study to Investigate the Feasibility, Reliability and Efficacy of Utilizing OCT to Automate PROSE Fitting

Keratoconus · Ocular Surface Disease · Pellucid Marginal Corneal Degeneration · Ectasia · Corneal Trauma

Enrolled (actual)
11
Serious AEs
0.0%
Results posted
Oct 2023
Primary outcome: Primary: Central Corneal Clearance — 354.29 microns

Study Design & Population

Study type
Interventional
Phase
N/A
Interventions
Ocular surface optical coherence tomography (Device)
Age
Adult, Older Adult · 21+ yrs
Sex
All
Sponsor
Boston Sight
Primary completion
Mar 2022

Outcome Measures

OutcomeResultp-value
PRIMARY
Central Corneal Clearance
354.29
PRIMARY
Lens Decentration
477.89
PRIMARY
Lens Comfort After 45 Minutes of Lens Wear
2.29
PRIMARY
Number of Participants With Lens Centration
5
PRIMARY
Number of Participants With Aligned Scleral Landing Zone
3
SECONDARY
Lens Rotation
28.83
SECONDARY
Number of Participants Achieving an Acceptable Fitting Lens
6
SECONDARY
Number of Participants With Lens Movement
4
SECONDARY
Number of Participants With Adequate Lens Limbal Clearance
7

Summary

EYEdeal Scanning Technology enables rapid measurement of corneal and scleral topography. By accurately measuring the ocular surface with this imaging technology, the current iterative fitting method required to fit PROSE (prosthetic replacement of the ocular surface ecosystem) devices could be replaced and/or strengthened with a more rapid, automated fitting sequence. This could considerably reduce the time needed per visit, the number of visits, and the number of devices needed to be manufactured to reach the endpoint. Additionally, the scanning technology may afford the opportunity to successfully fit some pathology that were previously treatment failures. The automated technology may as well reduce the intensive clinician training time needed to fit PROSE devices, therefore increasing the availability and access to patients. The goal of this research is to evaluate the reliability and efficacy of this automated technology for fitting PROSE devices. Data from real-time measurement of the human eye ocular surface topography will be used to fabricate a prosthetic lens. The fit of the PROSE device will be evaluated, as well as the subjective comfort of the fit.

Eligibility Criteria

Inclusion Criteria

  • Written Informed Consent has been obtained prior to any study-related procedures
  • Male or female, 18 years of age and older prior to the initial visit
  • Established wearer of PROSE in the study eye with an optimized fit in the opinion of the clinician
  • No medical need for a PROSE retreatment or replacement lens at the time of enrollment in the study eye, in the opinion of the clinician
  • Initial PROSE fitting was initiated and completed at BostonSight, Needham
  • Current PROSE device does not have channels or fenestrations
  • Currently wears a PROSE device in the study eye primarily for an irregular ocular surface /irregular cornea (including but not limited to Keratoconus, Pellucid Marginal Degeneration, Ectasia, Post-Penetrating Keratoplasty, Corneal Scar, Trauma) AND/OR ocular surface disease
  • Has the ability to NOT wear the current PROSE device in the study eye for 3 days prior to scan and 3 days prior to first fitting appointment in the opinion of the investigator and subject
  • In the opinion of the investigator, the subject has the ability to follow study instructions
  • In the opinion of the investigator, the subject has the ability to complete all study procedures and visits

Exclusion Criteria

  • Is currently participating in any other type of eye-related clinical or research study
  • Is pregnant or nursing as reported by the subject
  • Has a condition or is in a situation which, in the investigator's opinion, may put the subject at significant risk, may confound study outcomes, or may significantly interfere with the subject's participation in the study
  • Has had previous ocular surgery within the past 12 weeks
  • Intolerance to PROSE wear
  • Inability to maintain stable fixation and exposure for ocular surface imaging
  • Corneal touch by the posterior surface of the device in current PROSE device
  • Allergy to sodium fluorescein
  • Patient is an employee of BostonSight
  • Subject is currently incarcerated. -
View full record on ClinicalTrials.gov →