Mode
Text Size
Log in / Sign up
N/A Completed N=12 Other

SpO2 Accuracy of Noninvasive Pulse Oximeter Sensor

Healthy · Hypoxia
Source: ClinicalTrials.gov NCT05141500 ↗
Enrolled (actual)
12
Serious AEs
0.0%
Results posted
Dec 2022
Primary outcomePrimary: Accuracy of Sensor by Arms Calculation — 1.62; 1.56; 1.48; 1.72 Root Mean Square Error (%)

Summary

The purpose of this clinical study is to validate the SpO2 accuracy of the Stryker Sustainability Solutions pulse oximetry sensors during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. The end goal is to provide supporting documentation for the SpO2 accuracy validation of the reprocessed sensors.

Outcome Measures

OutcomeResultp-value
PRIMARY
Accuracy of Sensor by Arms Calculation
1.62; 1.56; 1.48; 1.72

Eligibility Criteria

Inclusion Criteria

  • Good general health with no evidence of pre-existing medical problems
  • Fluent in both written and spoken English
  • Must be able to review informed consent and is willing to comply with study procedures

Exclusion Criteria

  • Obesity (BMI >30)
  • Known history of heart disease, lung disease, kidney or liver disease
  • Prior diagnosis of asthma, sleep apnea, or use of CPAP
  • Diabetes
  • Clotting disorder
  • Hemoglobinopathy or history of anemia that in the opinion of the investigators would make them unsuitable for study participation
  • Any serious systemic illness
  • Current smoker
  • Any injury, deformity, or abnormality at the sensor sites that in the opinion of the investigator would interfere with the sensors working correctly
  • History of fainting or vasovagal response
  • History of sensitivity to local anesthesia
  • Prior diagnosis of Raynaud's disease
  • Unacceptable collateral circulation based on exam by investigators (Allen's test)
  • Pregnant, lactating, or trying to get pregnant
  • Unable or unwilling to provide informed consent, or unable or unwilling to comply with study procedures
  • Any other condition which in the opinion of the investigators would make them unsuitable for the study
View full record on ClinicalTrials.gov →

Data sourced from ClinicalTrials.gov (NCT05141500). Outcome figures and adverse-event rates are extracted automatically from the registry's posted results and are provided for clinician reference, not as a substitute for the primary publication. Informational only — not medical advice.

Back to search